Reducing Photic Phenomena and Retinal Background Illumination by Using an Intraocular Lens

Authors

DOI:

https://doi.org/10.20535/ibb.2020.4.4.214806

Keywords:

Intraocular lens, Crystalline lens, Gradient optics, Zemax, SolidWorks, Fresnel reflections

Abstract

Background. After implantation of monofocal intraocular lenses (IOLs), the risk of developing light phenomena is 9%, and after implantation of multifocal IOLs the one is 41%. These effects not only cause discomfort, but also poor image quality. The existing IOLs have a number of deficiencies that cause different types of photic phenomena.

Objective. The aim of the work is optical calculation and simulation of the parameters that an IOL should have in order to reduce photic phenomena and retinal background illumination while increasing the transmission contrast. We also are aimed to design a new IOL on the basis of the results obtained.

Methods. To calculate the reflected rays, we used the Snell–Descartes law, according to which we obtained the value of the critical angle for a hydrophobic acrylic IOL with the refractive index of 1.55 at the wavelength of λ = 0.55 μm. This is consistent with aqueous humor, the refractive index of which is 1.33. The calculation of the transmitted light loss was determined by Fresnel reflection coefficient. We handled with optical system in which the aperture diaphragm fulfilled the role of a pupil. Using the Zemax 13 software environment, we simulated ray path for a human eye, the sagittal axis of which is 28 mm at different height of IOL placement. We applied the results obtained to design a new IOL in the SolidWorks environment.

Results. The calculations made it possible to identify the shortcomings of modern IOLs and methods for their elimination. It was found that in order to reduce the risk of photic phenomena and, as a result, of increasing luminous transmission, an IOL should be placed at a distance of at least 4 mm from the iris. It should contain two or more optical layers, the refractive index of which changes towards the center of the lens, and have the surface roughness of 35 nm. Based on the calculations, we carried out simulation in the Zemax 13 environment, which confirmed their veracity. When simulated with these parameters, the standard deviation of an image fell completely within the Airy disk, which has a size of 3.598 μm with an image size of 2.972 μm. Thus, the optical system is considered diffraction limited and no further optical optimizations are possible. Using the Solidworks software and the results obtained, we proposed the proprietary IOL model called "NVision OP". This IOL has an optical part thickness of 1 mm with a diameter of 5 mm. In general, the hollow, volume-replacing IOL with a coating has a diameter of 10 mm and a thickness of 5 mm, the thickness of the coating is 150 μm.

Conclusions. The study revealed a number of factors that require improvement and elimination to prevent the occurrence of various types of photic effects. These include: lens surface roughness, IOL refractive power, shape, lens edge thickness, depth of IOL implantation into an eye, phacodonesis and lens displacement, aperture diaphragm diameter. After data optimization, according to the calculated results, we carried out the simulation in the Zemax 13 and Solidworks environments. On the basis of this simulation we proposed the model of an intraocular lens "NVision OP"; the photic effects namely arcs, flare, flashes, glare and halo are eliminated as much as possible. The hollow, volume-replacing IOL "NVision OP" has elements on its coat that allow to use the suture fixation, which prevents the dislocation of the IOL. For the implantation of the proposed IOL "NVision OP", it is recommended to use a viscoelastic and the Alcon injector with the cartridge B. Due to the fact that the shape of the IOL corresponds to the native human lens, the lens is located in the place of the phacoemulsified substance, and the implantation does not take much time.

References

Foster A. Cataract - a global perspective: Output, outcome and outlay. Eye. 1999;13(Pt 3b):449-53. DOI: 10.1038/eye.1999.120

Thylefors B, Negrel AD, Pararajasegaram R, Dadzie KY. Global data on blindness. Bulletin of the World Health Organization. 1995;73(1):115-21.

Thylefors B, Resnikoff S. [Progress in the control of world blindness and future perspectives]. Sante. 1998;8(2):140-3.

Takhchidi HP, Agafonova VV, Yanovskaya NP, Frankovska-Gerlak M. Efficiency of one-stage combined surgical treatment of cataract and open-angle glaucoma complicated by pseudoexfoliation syndrome. Ophthalmosurgery. 2008;(1):22-8.

Takhchidi HP, Egorova EV, Tolchinskaya AI. Intraocular correction in surgery of complicated cataracts. Moscow: New in Medicine, 2004, 169 p.

Kopaeva VG (ed.). Eye Diseases. Moscow: Medicine, 2002, 560 p.

Zhaboyedov DG. Method for diagnosing pseudophakia of the eye. Patent of Ukraine UA78758U, published 25.03.2013.

Davison JA. Positive and negative dysphotopsia in patients with acrylic intraocular lenses. Journal of Cataract and Refractive Surgery. 2000;26(9):1346-55. DOI: 10.1016/S0886-3350(00)00611-8

Haring G, Dick BH, Krummenauer F, Weissmantel U, Kröncke W. Subjective photic phenomena with refractive multifocal and monofocal intraocular lenses: Results of a multicenter questionnaire. Journal of Cataract and Refractive Surgery. 2001;27(2):245-9. DOI: 10.1016/S0886-3350(00)00540-X

Engren A-L, Behndig A. Anterior chamber depth, intraocular lens position, and refractive outcomes after cataract surgery. Journal of Cataract and Refractive Surgery. 2013;39(4):572-7. DOI: 10.1016/j.jcrs.2012.11.019

Cataract in the Adult Eye. San Francisco: American Academy of Ophthalmology, 2011.

Morozova TA. Intraocular correction of aphakia with a multifocal lens with gradient optics. Clinical and theoretical study: dissertation of a candidate of medical sciences. Moscow, 2006, 151 p.

Polishchuk OS, Kozyar VV. Flexible volume interchangeable multifocal intraocular lens "NVision OP". Patent of Ukraine UA142801U, published 25.06.2020.

Charman WN. Vision and Visual Dysfunction: Visual Optics and Instrumentation, vol. 1. Pan Macmillan, 1991.

Kolokolov AA. Fresnel formulas and the principle of causality. Physics-Uspekhi. 1999;42(9):931-40. DOI: 10.1070/PU1999v042n09ABEH000482

Born M, Wolf E. Fundamentals of Optics. Moscow: Nauka, 1973, 713 p.

Bennett HE, Porteus JO. Relation Between Surface Roughness and Specular Reflectance at Normal Incidence. Journal of the Optical Society of America. 1961;51(2):123-9. DOI: 10.1364/JOSA.51.000123

Khusu AP, Vitenberg YuR, Palmov VA. Surface roughness. A probability-theoretical approach. Мoscow: Nauka, 1975, 344 p.

Ovchinnikov SS, Tymkul VM, Kuznetsov MM. Optical Method for Monitoring Surface Roughness. Inter Expo Geo-Siberia. 2013;5(1):282-5.

Kuznetsov SL. Light Reflection from the Intraocular Lens and a Way to Reduce It. Theoretical Study. Ophthalmology in Russia. 2018;15(3):318-24. DOI: 10.18008/1816-5095-2018-3-318-324

Moskalev VA, Nagibina IM, Polushkina NA, Rudin VL. Applied Physical Optics. St. Petersburg: Politekhnika, 1995, 528 p.

Landsberg GS. Optics. Moscow: Fizmatlit, 2003, 848 p.

Knunyants IL (ed.). Chemical Encyclopedia. Vol. 2: Daf-Med. Moscow: Sovetskaya entsiklopediya, 1990, 671 p.

Zhaboyedov DG. Features of optical phenomena of natural and artificial crystalline lenses of a human eye. Problems of Environmental and Medical Genetics and Clinical Immunology. 2012;(5(113):529-53.

Gritsenko K. P. Polytetrafluoroethylene films deposited by evaporation in vacuum: growth mechanism, properties, application. Russian Chemical Journal. 2008;52(3):112-123.

Legeais J-M, Werner LP, Legeay G, Briat B, Renard G. In vivo study of a fluorocarbon polymer-coated intraocular lens in a rabbit model. Journal of Cataract and Refractive Surgery. 1998;24(3):371-9. DOI: 10.1016/S0886-3350(98)80326-X

Werner LP, Legeais J-M, Durand J, Savoldelli M, Legeay G, Renard G. Endothelial damage caused by uncoated and fluorocarbon-coated poly(methyl methacrylate) intraocular lenses. Journal of Cataract and Refractive Surgery. 1997;23(7):1013-9. DOI: 10.1016/S0886-3350(97)80073-9

Published

2020-12-31

How to Cite

1.
Polischuk A, Kozyar V, Zhaboedov D. Reducing Photic Phenomena and Retinal Background Illumination by Using an Intraocular Lens. Innov Biosyst Bioeng [Internet]. 2020Dec.31 [cited 2025Oct.10];4(4):199-210. Available from: https://ibb.kpi.ua/article/view/214806

Issue

Section

Articles